Mounting on a pipe stub, Mounting – Xylem PHIX COMPACT pH and Redox Transmitter User Manual
Page 12

12
Mounting
GB pHix Compact Manual 100318
Mounting on a pipe stub
In order to ensure a tight seal and correct
function of pHix
®
Compact, the following
should be complied with:
1: The pipe stub must be flush and free of
scales, burrs etc.
2: The pipe stub must have a length so
that the tip of the electrode is exposed
approximately 10 mm into the liquid.
3: Take care to have adequate room for
mounting and removal o f pHix
®
Com-
pact.
4: The working pressure of the system
must not exceed 5 bar g.
5: The working temperature of the s ystem
must not exceed 80 °C.
See also the following illustrations for correct
adaptation to the pipe system.
Approx. 10 mm
pHix
®
Compact is mounted correctly when the
electrode tip is exposed approximately 10 mm out
from the inner lining of the pipe.
Min. 25 mm
thread
2 " outer pipe
thread (ISO
228-1)
80 mm to
inner lining
Viton gasket
Min. 320 mm
free space for
mounting/re-
moval of pHix
®
Compact.
Max. working pressure = 5 bar g.
Maks. working temperature = 80 °C.