D. upgrade converter firmware – Xylem SuSix Turbidity and Suspended Solids Transmitter User Manual
Page 103
103
FW: 841013-005/843003-001
Turbidity and Suspended Solids Transmitter
EN 5.10 SuSix Manual 1406
D. Upgrade Converter Firmware
1. Select "Firmware" in the menu bar and click "Converter".
2. Locate the "MJK Field FW" directory on the CD-ROM, select the
"842xxx-xxx.hex" file (or similar) and click "Open".
3. Select the required SuSix
®
(option 1, 2, 3 or 4) and click OK when the
"Converter Firmware" dialogue box appears and prompts for selecting
the converter to program.
The converter firmware is then transferred, and the converter Flash
programming is initiated. Expect this process to take a few minutes.
4. When the Flash programming is done, click OK on the PC and the
SuSix
®
Display Unit to return to the SuSix
®
display.
5. If required you may adjust the SuSix
®
Display Unit screen contrast by
pressing the two outmost keys simultaneously, press the up/down
keys as required, and save the new setting by again pressing the two
outmost keys simultaneously.