The following symbols indicate the result is out of test range:

Result is higher than range

> >

Result is lower than range

< <

6 The ‘Options’ menu at the bottom then offers the choice to:

‘Choose a Test’ - return to the menu of test programs and select another test

‘Read’

- read further sample tubes of the currently selected test

‘Blank’

- re-blank the instrument

‘Follow-On’

- carry out a continuation test if available.


Continuation Tests (Certain Tests Only)

1 Select ‘Follow-On’ and press [OK] during the result display period of the

foregoing test stage. The ‘Insert Sample’ screen will appear.

Place

SAMPLE TUBE in the test chamber, then press [OK].

2 The instrument will take the reading and calculate the result from the

combination of readings (where appropriate). The result will be displayed:

Chlorine-Total / 5

Phot 008

1.50

mg/l Cl

2

Choose a Test Read Blank

Return

3 During the display period, the same options are available as at the end of a

normal test program. Select ‘Return’ from the ‘options menu' to take the
program back to the start of the first stage of a multiple test procedure to
enable further samples to be tested for the same parameters.

Note that some continuation test procedures involve a standing period. The
photometer may switch off automatically during this time. To avoid the
instrument switching off, use the timer function to time any standing period.
The timer will over-ride the auto switch off function.

Favorite Tests List

The four most recently used tests are listed at the top of the 'Choose a Test'
screen for convenience.

8