Troubleshooting, Specifications – YSI pH100A User Manual

Page 9

Advertising
background image

7

TROUBLESHOOTING

MAIN

DISPLAY

POSSIBLE CAUSE

CORRECTIVE ACTION

Er 1

 pH electrode offset is greater/less

than +/-1.5 pH.

 STAND was pressed before the

electrode and ATC/Temp probe settle
to within ±1.5 pH of the buffer value.

 pH electrode is faulty

 Replace the buffer and/or

the pH electrode. Press ESC.

 Allow sufficient time for the

electrode and ATC/Temp
probe to stabilize.

 Return for service.

Er 2

 pH electrode slope is off by more

than 30% of ideal slope.

 SLOPE was pressed before the

electrode and ATC/Temp probe
settled to within 30% of the buffer
value.

 Buffer 4.00, 4.01, 9.18 or 10.01 is not

correct.

 Check that the correct buffer

is used and that the
electrode slope is not off by
more than 30% from the
theoretical slope.

 Allow sufficient time for the

electrode and ATC/Temp
probe to stabilize.

 Replace the buffer and/or

the pH electrode. Press ESC.

 Return for service.

Er 3

 Temperature is out of the 0.0 to 60.0

°C range.

 Bring the buffer temperature

within range.

 Return for service.

OvEr/Undr

 Measured pH is out of the 16.00/-2.00

pH range.

 Measured mV is out of the 1250/-

2000 mV range.

 Measured temperature is out of the

-10/120 °C range.

 Bring the out of range unit

into the correct measuring
range.

 If units are within proper

range, return product for
service.


SPECIFICATIONS

Display Range Accuracy

Resolution

pH

-2.00 to 16.00 pH

±0.1%, ±2 lsd

0.01 pH

mV

-1999 to 1999 mV

±0.1%, ±1 lsd

1 mV

Temperature °C

-10.0 to 120 °C

±0.5 °C

0.1 °C

pH Temp Compensation

AUTO/MANual -10.0 to 120.0

°C

pH Buffer Recognition

(4.01, 7.00 & 10.01) or (4.00, 6.86 & 9.18)

pH Buffer Calibration Temp. Range

0 to 60

°C

pH Electrode Offset Recognition

±90mV at pH 7.00 or 6.86

pH Electrode Slope Recognition

±30% at pH 4.00, 4.01, 9.18 or 10.01

Power

One 9V battery/ Approximately 1000 hours

Instrument Casing

Waterproof IP 67

Calibration Back-up

Yes

Audio Feedback

Yes, on all touch keys

Autolock Feature

Yes

Operating Temp. Range

0 to 50

°C

Operating Relative Humidity Range

up to 95%

ATC Probe

Thermistor, 10k / 25

°C

Dimensions (L x W x D)

18.7 cm x 7.6 cm x 3.8 cm (7.37 in x 3 in x 1.5 in)

Weight (batteries included)

270 grams (.6 lb)