Wear parts and accessories

Cleaning

Remove water-soluble contamination by rinsing with deionized water. Re-

move other contamination as follows:

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household
washing-up liquid

Lime and hydroxide deposits Rinse with citric acid (10 % by weight)

Proteins

Immerse in pepsin cleaning solution PEP/pH

for approx. 1 hour.

Note: Make sure the level of the reference
electrolyte is above that of the cleaning

solution.

Note
Hydrofluoric acid, hot phosphoric acid and strong alkaline solutions destroy
the glass membrane.

After cleaning

Rinse the combination electrode with deionized water and condition it in

reference electrolyte solution (KCl 3 mol/l, Ag

+

free) for at least 1 hour. Then

recalibrate the combination electrode.

Wear parts and accessories

Description

Model

Order no.

Reference electrolyte solution 250 ml
(KCl 3 mol/l, Ag

+

free)

KCl-250

109 705Y