Sample-On-Demand: This feature is not supported by the PL200H data logger.

Alarm Times: Check the box

associated with each alarm to

enable that feature and enter

the desired start or stop time.

The logger will start at the next

occurrence of the Alarm 1 time and stop at the next occurrence of the Alarm 2 time

after the Alarm 1 timer has expired. Either or both alarms can be enabled.

Important:

These timers are a one time event; the logger will not restart on the next

occurrence of the Start timer unless the logger is reprogrammed with these

settings.

Use care when programming the alarm times. If left enabled, changing and

reprogramming any of the setup parameters will restart the alarm timers and

could cause the logger to stop recording and the loss of important data.

Using the Alarm 2 Stop Timer can significantly increase battery life, especially

when using short sample intervals. See section XVIII for more information on

battery life.