Xylem TB500 ONLINE TURBIDITY ANALYZERS User Manual

Page 32

Advertising
background image

TB500 (2/07)
Rev. 1.0

28

9.3 Diagnostic Chart

Symptom Cause Cure

Lower display shows MA

4-20 mA loop open

Check wiring. See sections
3.3.4 and 7.2

Lower display shows DESC

Desiccant pouch bad

Change desiccant pouch. See
section 10.2

Lower display shows LAMP

Lamp failed

Replace lamp. Refer to section
10.3

Lower display shows FLOW

Sample flow has stopped

Restore flow. Contact GW
about factory installed option

Lower display shows FAIL

Major system fault

Refer to section 9.1 & 9.2

Readings are higher than
expected

(1)

Bubbles in solution








(2)

Condensate or leaky
cuvette


(3)

Flow through cuvette
dirty

Instrument out of calibration

(1)

Ensure that the drain vent is

open and is not obstructed.
See section 3.2.2.
(2)

Apply backpressure. See

section 3.2 and figure 4

(3)

For sever cases of bubbles a

stilling chamber is
available. Call GW.

Part# 20106GW

Check flow through cuvette for

condensate or leaks.


Clean cuvette. See section 10.1


Recalibrate. Refer to section 5

Readings are erratic

(1)

Bubbles in solution

(2)

Debris in flow through

(1)

See above

(2)

Clean debris from cuvette

Readings are lower than expected Instrument out of calibration

Recalibrate. Refer to section 5

Upper display flashes

Sample Over-Range

Check sample. Sample may be
too high to read.


9.4 Technical and Customer Assistance
If for any reason assistance is needed regarding this instrument please do not hesitate to
contact Global Water’s Customer Service Department:

Global Water Instrumentation

11390 Amalgam Way

Gold River, CA 95670

Phone: (800) 876-1172

Fax: (847) 672-9988

Email:

[email protected]

www.globalw.com

Advertising