VII. Dissolved Oxygen Sensor

a. Dissolved Oxygen sensor specifications.

Output:

4-19mA

Range:

0-100% Dissolved Oxygen

Accuracy:

+/- 0.5% FS

Operating Voltage:

10-36VDC

Current Draw:

11.8 mA plus sensor output

Warm Up Time:

10 seconds minimum

Operating Temp:

-40° to +55°C

Size of Probe:

1 ¼” diameter x 11” long

Weight:

1 lb.

Membrane:

0.001 FEP Teflon (standard)

Combined Error:

2% FS

b. The sensing element is covered with a protective cap while at the

Global Water facility to prevent the sensor from becoming damaged.
This cap must be removed prior to sensor installation or the readings
will be in error. To remove the cap, remove the protective shield and

remove the rubber cap covering the sensing element. Replace the
protective shield. Note: Save the cap for future use.


c. This sensor has a removable sensing electrode. If the sensor is

reading incorrectly, after following the basic maintenance steps, the
electrode should be removed and the metal contacts cleaned. If this
does not improve the sensor’s results the electrode should be

replaced. To remove the sensing electrode loosen the set screw
holding the sensor shield and remove the shield from the sensor.
Unscrew the sensing electrode from the sensor housing. Clean the
metal contacts of the electrode with a clean cloth. If the electrode is

replaced with a new electrode the sensor must be recalibrated to work
correctly.

d. The Dissolved Oxygen sensor is a three-wire sensor. Three wire

sensors use the red wire for positive voltage, the white wire for the
output signal, and the black wire for ground. Warning: Always
connect the sensor with the power turned off.