Storage, Aging – Xylem Sentix ORP 900 User Manual

Page 6

Advertising
background image

Aging

SenTix

16

ba75845e01

08/2009

Storage

During short

measuring breaks

Immerse the electrode in the reference electrolyte with the refilling opening
open.

Prior to the next measurement, shortly rinse the electrode with the test
sample or deionized water.

Overnight or

longer

Insert the clean electrode into the watering cap filled with reference
electrolyte and shut the refilling opening.

Note
During longer storing periods, salt sediments may develop on the watering
cap. They do not affect the measuring characteristics and can easily be re-
moved with deionized water when the electrode is put into operation again.

Aging

Every ORP electrode undergoes a natural aging process. Extreme operating
conditions can considerably shorten the lifetime of the electrode. These are:

z

Strong acids or lyes, hydrofluoric acid, organic solvents, oils, fats, bro-
mides, sulfides, iodides, proteins

z

High temperatures

z

High changes in pH and temperature.

The warranty does not cover failure caused by measuring conditions and
mechanical damage.

Electrode

Reference electrolyte

WTW model
(see page 19)

SenTix

®

ORP 900

3 mol/l KCl, Ag

+

free

KCl-250 (250 ml)

Advertising