4 technical data, 1 general data, Technical data – Xylem TetraCon 925 User Manual
Page 6: General data, 4technical data

Technical data
TetraCon
®
925
14
ba75824e01
08/2009
4
Technical data
4.1
General data
General features
Measuring principle
Four-electrodes measurement
Cell constant
0.475 cm
-1
±1.5 %
Temperature sensor
Integrated NTC 30 (30 k
:
at 25 °C / 77 °F)
Dimensions
(in mm)
Weight
Approx. 60 g (without cable)
Materials
Shaft
Epoxy
Connection head
POM
Conductivity electrodes
Graphite
Thermistor enclosure
Graphite
Connection cable
Lengths
1,5 / 3 / 6 / 25 m
Diameter
4.3 mm
Smallest allowed
bend radius
Fixed installation:
20 mm
Flexible use:
60 mm
Plug type
Socket, 4 pins
Pressure
resistance
Sensor with connection
cable
IP 68 (2.5 x 10
5
Pa or 2.5 bar)
Cable plug
IP 67 (when plugged in)
The TetraCon
®
925 meets the requirements according to article 3(3) of the
directive, 97/23/EC ("pressure equipment directive").
120.0
176.0
15.3
21.7