ETS-Lindgren HI-2200 RF Survey Meter User Manual
Page 44

44 |
Download Utility
Download the File
When downloading new software to the HI-2200,
all logged data in the HI-2200 will be erased.
Before downloading new software to the
HI-2200, make sure to upload all logged data to
your computer.
Before downloading new software to the
HI-2200, make sure the battery is completely
charged.
1.
Press the function key under MISC on the main logging screen.
2.
Press the function key under DOWNLOAD, and then answer Yes at
the Are you sure? message.
3.
In Download Utility, select the correct communications port, 1 or 2. The
Progress area should display the message Device Found. You are
now ready to download new software.
To stop a download in progress, select the
Abort key on the computer screen. Aborting the
download process leaves the previous revision
of software on the HI-2200 running, but all data
is erased.
4.
Select the Download key and the software will download to the
HI-2200. When the download is complete, the HI-2200 will restart.