Measurement support functions – Atec Agilent-4287A User Manual
Page 13

DC resistance (Rdc) measurement
Trigger function
Averaging function
13
Measurement range
0.1
Ω to 100 Ω
Measurement resolution
1 m
Ω
Test signal level
1 mA (maximum)
Error correction
OPEN/SHOR T/LO AD Calibration, OPEN/SHORT
Compensation. (Changing the frequency or oscillator
level settings after the calibration or compensation
makes the calibration and compensation data invalid.)
Measurement uncertainty
[%]
R
dut
: DC resistance measurement v alue [
Ω]
(Within
± 5 °C from the calibration temperature.
Measurement accuracy applies when the calibration
is performed at 23
°C ± 5 °C. When the calibration is
performed beyond 23
°C ± 5 °C, the measurement
accuracy decreases to half that described.)
Trigger mode
Internal, External (external trigger input connector or
handler interface), Bus (GPIB or LAN), Manual (front key)
Setting range
1 to 100 (integer)
1
0.05
R
dut
---------
R
dut
10000
--------------
+
100
×
+
±
( )
Measurement Support Functions