Basic measurement characteristics – Atec Agilent-4287A User Manual
Page 6

q
[rad]
L, C, X, B
[%]
R, G
[%]
D
Q
D
x
=
Measurement value of D
Q
x
=
Measurement value of Q
E
a
=
(Within
± 5 °C from the calibration temperature.
Measurement accuracy applies when the calibration
is performed at 23
°C ± 5 °C. When the calibration is
performed beyond 23
°C ± 5 °C, the measurement
accuracy decreases to half that described.)
Oscillator level
≥ –33 dBm
Frequenc y
≥ 1 MHz,
≤ 100 MHz
± 0.65 [%]
Frequenc y > 100 MHz,
≤ 500 MHz
± 0.8 [%]
Frequenc y > 500 MHz,
≤ 1 GHz
± 1.2 [%]
Frequenc y > 1 GHz,
≤ 1.8 GHz
± 2.5 [%]
E
c
100
----------
±
E
a
E
b
+
(
)
2
E
c
D
x
(
)
2
+
±
E
a
E
b
+
(
)
2
E
c
Q
x
(
)
2
+
±
D
x
E
c
100
----------
tan
1
<
1
D
x
2
+
(
)
E
c
100
----------
tan
1
D
x
±
E
c
100
----------
tan
-------------------------------------------------------
±
D
x
0.1
≤
E
c
100
----------
±
Q
x
E
c
100
----------
tan
1
<
1
Q
x
2
+
(
)
E
c
100
----------
tan
1
Q
x
±
E
c
100
----------
tan
-------------------------------------------------------
±
10
E
c
------
Q
x
10
≥
≥
Q
x
2
E
c
100
----------
±
( )
( )
( )
( )
( )
( )
6
Basic Measurement Characteristics
Definition of each parameter