Specifications – Atec AEMC-3710-3730 User Manual

Page 4

Advertising
background image

Specifications

Chauvin Arnoux, Inc. d.b.a. AEMC

®

Instruments · 99 Chauncy St. · Boston, MA 02111 USA · (800) 343-1391 · (617) 451-0227 · Fax (617) 423-2952

Visit our website at www.aemc.com

Call the AEMC

®

Instruments Technical Assistance Hotline for immediate consultation with an applications engineer: (800) 343-1391

950.BR - 3710/30 05/99

ORDERING INFORMATION

CATALOG NO.

Clamp-on Ground Resistance Tester Model 3710. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #1221.01
includes hard carrying case, 9V Alkaline battery, 25

calibration check loop, and user manual

Clamp-on Ground Resistance Tester Model 3730. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #1221.02
includes hard carrying case, 9V Alkaline battery, 25

calibration check loop, and user manual

Operating Humidity: 0 to 90% RH @
14°F to 104°F (-10°C to 40°C),
75% RH @ 131°F (55°C)

Storage Temperature:
-40°F to 158°F (-40°C to 70°C)

LCD: 3 3/4 Digit, 1.73" x 1.10" (44 x 28 mm)

U.S. Patent: No. 362,639

NSN: 6625-01-377-8030

Measurement

Range

Range

1.0 to 50.0

50.0 to 100.0

100 to 200

200 to 400

400 to 600

600 to 1200

1 to 300mA

0.300A to 3.000A

3.00A to 30.00A

Resolution

0.1

0.5

1

5

10

50

1mA

0.001A

0.01A

Accuracy*

± (1.5% + 0.1

)

± (2.0% + 0.5

)

± (3.0% + 1

)

± (6.0% + 5

)

± (10% + 10

)

n/a

± (2.5% + 2mA)

± (2.5% + 2mA)

± (2.5% + 20mA)

GROUND OR LEAKAGE CURRENT

GROUND RESISTANCE

Carrying Case (included)

* Reference conditions: 23°C ± 3K, 50% RH ± 10%, battery at 8V ± 0.2V, external magnetic field
< 40 A/m, external electrical field < 1 V/m, conductor centered, loop resistance noninductive. Accuracy % of
reading.

ELECTRICAL

Resistance Measurement Frequency:
1.689kHz

Current Measurement Frequency:
47 to 800Hz

Current Overload:
OL displayed above 30Arms

Power Supply: 9V Alkaline battery
(IEC 6LF22 or NEDA 1604A)

Battery Life: typical: 8 hours or approx.
1,000 measurements of 30 seconds

SAFETY

UL, CSA, GS Listed

IEC 1010-2-032 Double Insulation

Environmental: IP30, IEC 359 Group III

Vibration Test: IEC 68-2-6

Shock Test: IEC 68-2-27

Drop Test (1m): IEC 68-2-32

Dielectric Test: 2500V

AC

Working Voltage:
IEC1010 150V, Cat. III - Pollution Degree 2
IEC1010 300V, Cat. III - Pollution Degree 1

Max Overload (A or

Function):

100A continuous, 200A (<5s) 50/60Hz

MECHANICAL

Dimensions: 9.25" x 3.94" x 2.17"
(235 mm x 100 mm x 55 mm)

Weight: 2.2 lbs. (1 kg)

Case Material: Lexan

®

920A (UL94V2)

Jaw Cover Material: Lexan

®

500R with

10% fiberglass charge (UL94V0)

LCD Cover Material:
Lexan

®

920A (UL94V1)

Color: Gray body, red jaws

Jaw Window Diameter: 1.25" (32 mm)

Jaw Opening: 1.38" (35 mm)

Operating Temperature:
14°F to 131°F (-10° to 55°C)

Calibration Check Loop (included)

Autoranging

1.0 to 1200

Autoranging

1mA to 30.00

Arms

Advertising