DATA SHEET > CWS 500N4 > 20121113

TECHNICAL DETAILS

TESTING AS PER IEC/EN 61000-4-16

OUTPUT CHARACTERISTICS CONTINUOUS MODE

Test frequencies

DC, 16 2/3Hz, 50Hz and 60Hz

Signal level

0.1 - 35Vrms or DC

Test level 1

1V continuous

Test level 2

3V continuous

Test level 3

10V continuous

Test level 4

30V continuous

Output impedance

50ohm ± 10%

Total harmonics
distortion

< 10% (sinusoidal waveform)

Ripple on DC

< 5%

OUTPUT CHARACTERISTICS SHORT-TERM MODE

Test frequencies

DC, 16 2/3Hz, 50Hz and 60Hz

Signal level

1 - 330Vrms or DC

Test level 1

10V for 1s duration

Test level 2

30V for 1s duration

Test level 3

100V for 1s duration

Test level 4

300V for 1s duration

Output impedance

50ohm ± 10%

Total harmonics
distortion

< 10% (sinusoidal waveform)

Phase
synchronisation

0° ± 5%

Ripple on DC

< 5%

Fall/rise time

between 1us and 5us

Note:

These tests require either the
optional MV 2606N2.2 motor variac
or the optional programmable AC
voltage source ACS 500N2.3.

TESTING AS PER IEC/EN 61000-4-16

OUTPUT CHARACTERISTICS 15HZ - 165KHZ

Frequency range

10Hz to 165kHz

Signal level

0.1 to 35Vrms

Test level 1

1V - 0.1V - 1V

Test level 2

3V - 0.3V - 3V

Test level 3

10V - 1V - 10V

Test level 4

30V - 3V - 30V

Output impedance

50ohm ± 10%

Total harmonics
distortion

< 1% (sinusoidal waveform)

MEASUREMENT

Voltage

internal rms measurement

TEST ROUTINES

Quick Start

Immediate start; easy to use and fast
discrete frequency testing

Service

Service, Set-up

TESTING OF ELECTRICITY METERING
EQUIPMENT AS PER TR-EN 50579

INTERFERENCES DISTURBANCES TESTING AS PER TR-EN
50579 AND IEC 61000-4-19 ANNEX C (OPTION)

Test frequencies

2kHz to 150kHz

Signal level

Max. 3A up to 150kHz

Total harmonic
distortion (THD)

<5% at maximum level

Accuracy

better than ± 5%

Frequency step

Max. 1%

Output impedance

1.0ohm (external decoupling
impedance SH 1R)

www.emtest.com