System amplitude accuracy, Linear to log scale switching uncertainty, Display scale fidelity – Atec Agilent-8542E-8546A User Manual

Page 2: Gain compression, Third order intercept point

Advertising
background image

Amplitude Specifications

Characteristic Noise Indication with CISPR Measurement Bands

(0 dB attenuation, 50

input termination)

Band A, 9 to 150 kHz (200 Hz BW)

Peak

Quasi-Peak

Average

Preamp off

15 to –15 dBµV

6 to –25 dBµV

3 to –27 dBµV

Preamp on

2 to –28 dBµV

–7 to –29 dBµV

–9 to –31 dBµV

Band B, 150 kHz to 30 MHz (9 kHz BW)

Preamp off

–3 dBµV

–11 dBµV

–18 dBµV

Preamp on

–8 dBµV

–15 dBµV

–21 dBµV

Band C, 30 MHz to 1 GHz (120 kHz BW)

Preamp off

9 dBµV

2 dBµV

–5 dBµV

Preamp on

4 dBµV

–2 dBµV

–10 dBµV

System Amplitude Accuracy

Band 1

Band 2

Band 3*

9 kHz to 50 MHz

20 MHz to 2.9 GHz

1 to 6.5 GHz

Specification

± 2 dB

± 2 dB

Characteristic

±1 dB

±1 dB

± 3 dB

Linear to Log Scale Switching Uncertainty

85422E/85462A

± 0.25 dB at reference level

Display Scale Fidelity

85422E/85462A

Log maximum cumulative

(0 to –66 dB from reference level, 0 to –64 dB for Band 3 only)

3 kHz to 3 MHz IF BW

±(0.3 dB + 0.01 x dB from reference level)

1 kHz IF BW

±(0.4 dB + 0.01 x dB from reference level)

Log incremental accuracy

±0.4 dB/4 dB
(0 to –56 dB from reference level; 0 to –54 dB for Band 3 only )

Linear scale

±3% of reference level

Gain Compression

(Specification is derived from measured distortion with a total power at the input mixer of

10 dBm.

If the IF BW

300 Hz, this applies only if signal separation

4 kHz and the signal amplitude is

reference level + 10 dB.)

Band 1

Band 2

Band 3*

9 kHz to 50 MHz

20 MHz to 2.9 GHz

1 to 6.5 GHz

200 kHz

f

0

<10 MHz

< 0.75 dB

< 0.75 dB

< 0.75 dB

f

0

10 MHz

< 0.5 dB

< 0.5 dB

< 0.5 dB

Characteristic 1 dB compression point

8542E/8546A

( f

0

10 MHz)

Preamp off

89 dBµV

89 dBµV

102 dBµV

Preamp on

77 dBµV

77 dBµV

77 dBµV

(9 kHz <f

0

<

10 MHz)

Preamp off

85 dBµV

Preamp on

72 dBµV

85422E/85462A

( f

0

>

10 MHz)

(No bands)

Preamp off

102 dBµV

Preamp on

75 dBµV

(9 kHz

f

0

10 MHz)

Preamp off

95 dBµV

Preamp on

68 dBµV

Third Order Intercept Point

Band 1

Band 2

Band 3*

Bypass

f

0

> 200 kHz, signal separation >50 kHz

9 kHz to

20 MHz to

1 to

9 kHz to

8542E/8546A

50 MHz

2.9 GHz

6.5 GHz

2.9 GHz

Preamp off

97 dBµV

97 dBµV

112 dBµV

112 dBµV

Preamp on

85 dBµV

85 dBµV

85 dBµV

85 dBµV

85422E/85462A

(No Bands)

Preamp off

112 dBµV

Preamp on

85 dBµV

* For 8546A EMI receiver only

Advertising