Measurement throughput summary, Number of points – Atec Agilent-E5061B User Manual
Page 29

29
Measurement throughput summary
Table 36. Typical cycle time for measurement completion
1, 2
(ms) (Display update: off)
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 30 kHz IF bandwidth
Uncorrected
5
15
26
89
2-port cal
14
33
56
181
Start 1 GHz, stop 1.2 GHz, 300 kHz IF bandwidth
Uncorrected
4
9
15
43
2-port cal
11
21
33
88
Start 1 MHz, stop 3 GHz, 30 kHz IF bandwidth
Uncorrected
10
23
37
119
2-port cal
24
48
78
241
Start 1 MHz, stop 3 GHz, 300 kHz IF bandwidth
Uncorrected
9
17
26
73
2-port cal
20
37
54
148
Table 37. Typical cycle time for measurement completion
1
(ms) (Display update: on)
Number of points
51
201
401
1601
Start 1 GHz, stop 1.2 GHz, 30 kHz IF bandwidth
Uncorrected
45
47
49
103
2-port cal
55
59
69
195
Start 1 GHz, stop 1.2 GHz, 300 kHz IF bandwidth
Uncorrected
45
47
50
64
2-port cal
55
59
64
103
Start 1 MHz, stop 3 GHz, 30 kHz IF bandwidth
Uncorrected
45
47
50
133
2-port cal
55
61
90
255
Start 1 MHz, stop 3 GHz, 300 kHz IF bandwidth
Uncorrected
45
47
50
87
2-port cal
55
59
67
163
1. Typical performance.
2. Measured with the fi rmware revision A.02.00.