Number of points – Atec Agilent-E5061B User Manual
Page 30
Advertising
30
1. Typical performance.
2. Measured using a VEE Pro 9.0 program running on a 2.4 GHz Pentium
®
4,
Transferred complex S
11
data, using :CALC:DATA?SDATA.
3. Measured using E5061B VBA macro running inside the analyzer. Transferred complex S
11
data.
Table 38. Data transfer time
1
(ms)
Number of points
51
201
401
1601
SCPI over GPIB
2
REAL 64
6
16
29
112
ASCII
29
109
215
848
SCPI over GPIB/USB (82357B)
REAL 64
8
14
23
77
ASCII
73
282
563
2247
SCPI over 100 Mbps LAN (SICL-LAN)
2
REAL 64
5
5
6
8
ASCII
4
8
13
41
SCPI over 100 Mbps LAN (Socket)
2
REAL 64
2
2
3
4
ASCII
20
72
142
562
SCPI over USB
2
REAL 64
3
3
4
5
ASCII
4
10
19
69
COM (program executed in the analyzer)
3
Variant type
1
1
1
1
Advertising