13 measurement examples – Atec Anritsu-ME7838A User Manual
Page 10

10
PN: 11410-00593 Rev. D
ME7838A BB/mm-Wave VNA TDS
Broadband Specifications and Performance
VectorStar ME7838A BB/mm-Wave VNA
2.13 Measurement Examples
The following figures are measurement examples of the ME7838A Broadband system performance.
10 GHz
110 GHz
Dynamic range of ME7838A system at the W1 1 mm coaxial test port from 70 kHz to 125 GHz.
Pow er Meter vs a1/1 Receiver
–10.5
–10.4
–10.3
–10.2
–10.1
–10
–9.9
–9.8
–9.7
–9.6
–9.5
0
10
20
29
39
48
58
67
76
86
95
105
114
125
Frequency (GHz)
dB
m
|a1/1|
Power Meter
An example of power measurement agreement: power sensor vs. ME7838A a1 reference receiver.
Power sweep range at 77 GHz. By using detection and power control
inside the 3743A millimeter-wave module; improved accuracy,
linearity and range can be achieved.
Power sweep range at 94 GHz demonstrating
greater than 50 dB of control.