Maximum input voltages and channel isolation

Figure 5. Maximum safety voltage for scope reference to earth

Figure 6. Maximum input voltage

U1610A and U1620A

Maximum input voltages

Input CH1 and CH2 direct (1:1 probe)

300 Vrms CAT III

Input CH1 and CH2 (1:10 probe)

600 Vrms

1

CAT III, 1000 Vrms

1

CAT II

Input CH1 and CH2 (1:100 probe)

600 Vrms

1

CAT II, 1000 Vrms

1

CAT II, 3540 Vrms

1

CAT I

Meter input

600 Vrms CAT III, 1000 Vrms CAT II

Scope input

300 Vrms CAT III

Voltage ratings

Vrms 50–60 Hz (AC sine wave), VDC (DC applications)

Channel isolation

From any terminal to earth ground

600 Vrms CAT III

CAT ll 1000

1000

100

10

1

0.01 0.1 1 10 100

CAT lll 600

1000

100

10

1

0.0001 0.001 0.01

0.1

1

10

100 1000

Max input voltage

1. Refer to the respective probe's manual for more information on the specification