Digital multimeter specifications (continued), Data logger specifications, Scope and meter logger – Atec Agilent-U1610A-U1620A User Manual

Page 9

Advertising
background image

9

Digital multimeter specifications (continued)

Maximum
reading

10,000 counts with automatic polarity indication

Voltage

1

CAT II 1000 V or CAT III 600 V

Function

Range

Resolution

Accuracy

Input impedance

(nominal)

Test current

Temperature

3

-50 to 1000 °C

1 mV/°C

-50 to -21 °C

2.5% + 2 °C

7

-20 to 350 °C

0.5% + 2 °C

7

351 to 500 °C

1.75% + 2 °C

7

501 to 1000 °C

2% + 2 °C

7

-58 to 1832 °F

1 mV/°F

-58 to -5.8 °F

2.5% + 3.6 °F

7

-4 to 662 °F

0.5% + 3.6 °F

7

664 to 932 °F

1.75% + 3.6 °F

7

933 to 1832 °F

2% + 3.6 °F

7

Frequency

3

100.00 Hz

0.01 Hz

0.03% + 3

1000.0 Hz

0.1 Hz

10.000 kHz

0.001 kHz

100.00 kHz

0.01 kHz

1000.0 kHz

0.1 kHz

1. Only allowed to measure up to CAT III 600 V if referring to GND.

2. Only allowed for floating voltage.

3. Denotes typical specifications, all others are warranted.

4. The accuracy is specified after the Null function is used to subtract the test lead resistance and thermal effect.

5. RH is specified for < 60%. The temperature coefficient is 0.15 × specified accuracy as > 50 MΩ.

6. The accuracy is based on film capacitors or better and uses the Relative mode for residual values.

7. The accuracy is based on using the Null function to reduce the thermal effect.

8. Denotes

characteristics.

Data logger specifications

Scope and meter logger

Range

1 s/div – 86400 s/div (1 day/div)

Recording time span

8 days

Memory depth

691200 points

Recording mode

Continuous (Range will change according to the time elapsed)

Sampling rate

1 sample/s

Advertising