Single-output: 200 w, Continued) – Atec Agilent-6544A User Manual

Page 2

Advertising
background image

32

Supplemental Characteristics
for all model numbers

dc Floating Voltage: Output terminals can
be floated up to ±240 Vdc from chassis

ground

Remote Sensing: Up to half the rated
output voltage can be dropped in each

load lead. The drop in the load leads

subtracts from the voltage available for

the load.

Output Programming Response Time: The
rise and fall time (10/90% and 90/10%)

of the output voltage is less than 15 ms.

The output voltage change settles within

1 LSB (0.025% x rated voltage) of

final value in less than 60 ms.

Down Programming: An active down
programmer sinks approximately 20% of

the rated output current

Modulation: (Analog programming of
output voltage and current)
Input Signal: 0 to –5 V
Input Impedance: 10 k Ohm nominal

ac Input: (ac input frequency 47 to 63 Hz)

Voltage

100 Vac 120 Vac 220 Vac 240 Vac

Current

4.4 A

3.8 A

2.2 A

2.0 A

Input Power: 480 VA, 400 W at full load;
60 W at no load
Regulatory Compliance: Conforms to
UL1244 and IEC 61010-1.

Size: 425.5 mm W x 88.1 mm H x
439 mm D (16.75 in x 3.5 in x 17.3 in)

See page 101 for more details

Weight: Net, 14.2 kg (31.4 lb); shipping,
16.3 kg (36 lb)

Warranty Period: Three years

Single-Output: 200 W

(Continued)

S p e c i f i c a t i o n s

(at 0

˚

to 55

˚

C unless

otherwise specified)

Power Products Catalog 2002-2003

For more detailed specifications see the product manual at

www.agilent.com/find/power

Agilent Performance dc Power Supplies

speed and accuracy for test optimization

6541A-

6544A-

6545A-

J04

J09

J05

Special Order

Special Order

Special Order

Option

Option

Option

Number of outputs

1

1

1

GPIB

No

No

No

Output ratings

Output voltage

13 V

70 V

150 V

Output current (40° C)

15.3 A

3 A

1.2 A

Maximum current (50° C/55° C)

13.77 A/13 A

2.7 A/2.55 A

1.08 A/1.02 A

Programming accuracy at 25

˚

C ±5

˚

C

Voltage

0.06% +

8.5 mV

31 mV

65 mV

Current

0.15% +

21 mA

4.1 mA

1.7 mA

Ripple and noise

from 20 Hz to 20 MHz

Voltage rms

300 µV

600 µV

900 µV

peak-peak

3 mV

6 mV

9 mV

Current rms

8 mA

1.5 mA

1 mA

Load regulation

Voltage

1 mV

4.5 mV

7 mV

Current

1 mA

0.25 mA

0.25 mA

Line regulation

Voltage

0.5 mV

1.5 mV

2.5 mV

Current

1 mA

0.25 mA

0.25 mA

Transient response time

Less than 100 µs for the output voltage to recover to its previous level
(within 0.1% of the voltage rating of the supply or 20 mV, whichever is greater)
following any step change in load current of up to 50% of rated current

Supplemental Characteristics

(Non-warranted characteristics determined by design and
useful in applying the product)

Average resolution

Voltage

3.5 mV

1.4 mV

37.5 mV

Current

5 mA

1.2 mA

0.5 mA

OVP

23 mV

110 mV

250 mV

OVP accuracy

260 mV

1.5 mV

3 V

Ordering Information

Opt 100 87 to 106 Vac, 47 to 63 Hz

Opt 120 104 to 127 Vac, 47 to 63 Hz

Opt 220 191 to 233 Vac, 47 to 63 Hz

Opt 240 209 to 250 Vac, 47 to 63 Hz

* Opt 908 Rack-mount Kit (p/n 5063-9212)

* Opt 909 Rack-mount Kit w/ Handles

(p/n 5063-9219)

Opt 0L2 Extra Standard
Documentation Package

Opt 0B3 Service Manual

Opt 0B0 No documentation package

* Support rails required

Accessories

p/n 1494-0060 Accessory Slide Kit

E3663AC Support rails for Agilent
rack cabinets

Advertising