Atec California-Instruments-6000LS User Manual

Page 4

Advertising
background image

Measurements

* Accuracy specifications are in % of reading. Frequency measurement accuracy applies for output voltages of 20Vrms and
higher.

All specifications same as standard Ls models except where noted.

300 V rms output to chassis
Output Relay
Push button controlled and bus controlled output relay

Parameter

Range

Accuracy* (±)

Resolution

1 ø mode

3 ø mode

Frequency

45.00 - 81.91 Hz
82.0 - 819.1 Hz
> 819 Hz

0.1% + 1 digit

0.01 Hz
0.1 Hz
1 Hz

Phase

45 - 100 Hz
100 - 1000 Hz

0.5°

00.1°

Voltage (AC)

0 -300 V

0.05% + 250mV

10 mV

Current (AC rms)

0 - 50 A

0.1% + 150 mA

0.1% + 50 mA

1 mA

Real Power

0 - 5 kW

0.15% + 9 W

0.15% + 3 W

1 W

Apparent Power

0 - 5 kVA

0.15% + 9 VA

0.15% + 3 VA

1 VA

Power Factor

0.00 - 1.00

0.03

0.01

0.01

System

Setup storage
16 complete instrument setups
Transient lists storage
100 transient steps per list (SCPI mode) or 16 transient
registers (APE mode).

Trigger input
Triggers measurements or transient steps
SMA connector: 10K pull-up

Trigger output
SMA connector: HC TTL output.

Remote Control

Ethernet Interface
Local Area Network 10BaseT, 100BaseT, RJ45 connector.
(Option -LAN)
IEEE-488 Interface
IEEE-488 (GPIB) talker listener.
Subset:
AH1, DC1, DT1, L3, RL2, SH1, SR1, T6
RS232C Interface

Code and Format
SCPI
APE (option -GPIB)

Bi-directional serial interface.
9 pin D-shell connector
Handshake: CTS, RTS
Data bits:

7 with parity,
8 without parity

Stop bits:

2

Baud rate:

9600 to 115200

Supplied with RS232C cable

Physical

Dimensions (per chassis)
Height: 10.5" (267 mm)
Width: 19" (483 mm)
Depth: 23.7" (602 mm)
Depth includes rear panel connectors and safety covers.
All weights and dimension are per chassis.

Weight (per chassis)
193 lbs / 87.7 Kg net.
280 lbs / 127.3 Kg shipping
Vibration and Shock
Designed to meet NSTA project 1A transportation levels
Air Intake/Exhaust
Forced air cooling, side air intake, rear exhaust
Operating Temperature
0 to 35° C, full power.
Storage Temperature
-40 to 85° C.
Diagnostics
Built-in self test available over bus. (*TST).
Rear Panel Connectors
Three phase AC Input terminal block with safety cover
Three phase AC Output terminal block with safety cover
IEEE-488 (GPIB) connector (Option -GPIB)
9 pin D-Shell RS232C connector
Remote Inhibit (INH) and Discrete fault Indicator (DFI)
Remote voltage sense terminal block
Trigger In1 and Trigger Out1
System Interface connectors
Aux. Output (Option -AX)
(*RS232 DB9 to DB9 cable supplied)

Option Specifications

Advertising