Relative level accuracy measures the accuracy of a step change from
any power level to any other power level. This is useful for large
changes (i.e. 5 dB steps).
Repeatability measures the ability of the instrument to return to a given
power setting after a random excursion to any other frequency and power
setting. It should not be confused with absolute level accuracy.
Relative level accuracy at 850 MHz initial power +10 dBm
Final power (dBm)
Power error (dB)
0
–20
–40
-60
–80
–100
–120
–140
0.5
0.4
0.3
0.2
0.1
0
–0.1
–0.2
–0.3
–0.4
–0.5
Upper std dev (1 σ)
Mean
Lower std dev (1 σ)
ALC linearity 1900 MHz, CW, relative to 0 dBm
Amplitude (dB)
–20
–15
–10
–5
0
5
10
0.3
0.2
0.1
0
–0.1
–0.2
–0.3
Error (dB)
Upper std dev
(1 σ)
Mean
Lower std dev
(1 σ )
ALC linearity 850 MHz, CW, relative to 0 dBm
Amplitude (dB)
–20
–15
–10
–5
0
5
10
0.3
0.2
0.1
0
–0.1
–0.2
–0.3
Error (dB)
Upper std dev
(1 σ)
Mean
Lower std dev
(1 σ)