Inputs/outputs, agilent 8560 ec series, Options, Continued – Atec Agilent-8560EC-65EC Series User Manual

Page 9

Advertising
background image

Inputs/Outputs, Agilent 8560 EC Series,

continued

Blanking/gate

Output

Shared

BNC

female,

50

Ω

Blanking mode

During sweep

Low TTL level

During retrace

High TTL level

Gate mode

Gate on

High TTL level

Gate off

Low TTL level

External/gate

Trigger input

Shared BNC female, > 10 kΩ

Settable to high TTL or low TTL

GPIB

IEEE-488

bus

connector

Interface functions

SH1, AH1, T6, L4, LE0, RL1, PP1, DC1, DT1, C1, C28, TE0, SR1

Direct printer output

Supports HP 3630A PaintJet printer, HP 2225A ThinkJet printer

Direct plotter output

Supports HP 7225A/7440A/7470A/7475A/7550A

Options

Option 856xEC-001 second IF output, Agilent 8560 EC Series

(all values are nominal)

3 dB bandwidth

8560EC

8561EC

8562EC

8563EC

8564EC, 8565EC

NF

conversion gain

30 Hz to 2.9 GHz

1

> 25 MHz

> 25 MHz

> 25 dB

> 25 MHz

> 25 MHz

24

dB

25

dB

20

dB

25

dB

28

dB

1.2

dB

–6.5

dB

–1.2

dB

–1.2

dB

–1.2

dB

2.9 GHz to 6.5 GHz

> 30 MHz

> 30 MHz

> 30 MHz

> 30 MHz

26

dB

22

dB

22

dB

23

dB

–1

dB

–3

dB

–1

dB

–1

dB

6.5 GHz to 13.2 GHz

> 37 MHz

> 37 MHz

> 37 MHz

26

dB

26

dB

28

dB

–5.7

dB

–5.7

dB

–5.7

dB

13.2 GHz to 22 GHz

> 45 MHz

> 45 MHz

30

dB

32

dB

–8

dB

–8

dB

22 GHz to 26.8 GHz

> 45 MHz

> 45 MHz

32

dB

35

dB

–8

dB

–8

dB

26.8

GHz

to

31.15

GHz

>

25

MHz

28

dB

–9

dB

31.15

GHz

to

40

GHz

>

25

MHz

38

dB

–19

dB

40

GHz

to

50

GHz

>

25

MHz

42

dB

–23

dB

Option 8560EC-002 Built-in tracking generator

2

Frequency specifications

Frequency range

300 kHz to 2.9 GHz

Accuracy

After peaking

±(frequency reference accuracy x tuned frequency + 5% x span + 295 Hz)

Tracking drift (nominal)

Usable in 1 kHz RBW after 5 minutes warm-up. Usable in 300 Hz RBW after

30-minute

warm-up.

Minimum RBW

300

Hz

3

9

1. DC coupled for frequencies below 100 kHz. Option 856xEC-006 required for operation below 9 kHz in Agilent 8563EC, 8564EC, 8565EC.

2. Option 8560EC-002 is no longer offered.

3. Tracking generator not usable with resolution bandwidths ≤ 100 Hz.

Advertising