Atec Agilent-N9010A User Manual
Page 13

13
Phase noise
1
Offset
Specification
Typical
RF/MW
(Option 503, 507, 513, 526)
Noise sidebands
(20 to 30 °C, CF = 1 GHz)
100 Hz
–84 dBc/Hz
–88 dBc/Hz
1 kHz
–
–98 dBc/Hz nominal
10 kHz
–103 dBc/Hz
–105 dBc/Hz
100 kHz
–115 dBc/Hz
–117 dBc/Hz
1 MHz
–135 dBc/Hz
–137 dBc/Hz
10 MHz
–
–148 dBc/Hz nominal
1. Apply for all the RF/MW instruments with Serial number prefix ≥ MY/SG/US5340. Those instruments ship standard with N9010A-EP3 as the
identifier. For nominal phase noise values with the RF/MW EXA at other center frequencies refer to Figure 3. For earlier instruments refer to the EXA
specifications guide.
Figure 3. Nominal phase noise at different center frequencies for RF/MW EXA (Option 503, 507, 513, or 526) with SN prefix >/= MY/SG/US5340 that
ships standard with N9010A-EP3.
Nominal phase noise at different center frequencies for RF/μW EXA
with RBW selectivity curves, optimized phase noise versus offset frequency