Amplitude Accuracy and Range Specifications
Amplitude range
Measurement range
Displayed average noise level (DANL) to +23 dBm
Input attenuator range
(10 Hz to 44 GHz)
Standard
Option FSA
0 to 60 dB in 10 dB steps
0 to 60 dB in 2 dB steps
Electronic attenuator (Option EA3)
Frequency range
10 Hz to 3.6 GHz
Attenuation range
Electronic attenuator range
Full attenuation range
(mechanical + electronic)
0 to 24 dB, 1 dB steps
0 to 84 dB, 1 dB steps
Maximum safe input level
Average total power
(with and without preamp)
+30 dBm (1 W)
Peak pulse power
< 10 μs pulse width, < 1 % duty cycle +50 dBm (100 W) and input attenuation ≥ 30 dB
DC volts
DC coupled
AC coupled
± 0.2 Vdc
± 100 Vdc
Display range
Log scale
0.1 to 1 dB/division in 0.1 dB steps
1 to 20 dB/division in 1 dB steps (10 display divisions)
Linear scale
10 divisions
Scale units
dBm, dBmV, dBμV, dBmA, dBμA, V, W, A