Frequency specifi cations – Atec Agilent-E4443A-45A-40A-47A-46A-48A User Manual

Page 4

Advertising
background image

4

Frequency Specifi cations

Frequency range

E4443A

(DC coupled)

3 Hz to 6.7 GHz

(AC coupled)

20 MHz to 6.7 GHz

E4445A

(DC coupled)

3 Hz to 13.2 GHz

(AC coupled)

20 MHz to 13.2 GHz

E4440A

(DC coupled)

3 Hz to 26.5 GHz

1

(AC coupled)

20 MHz to 26.5 GHz

1

E4447A

(DC coupled)

3 Hz to 42.98 GHz

E4446A

(DC coupled)

3 Hz to 44 GHz

1

E4448A

(DC coupled)

3 Hz to 50 GHz

1

1. 325 GHz with external mixers

Band

Harmonic mixing mode (N)

0

1–

3 Hz to 3 GHz

1

1–

2.85 GHz to 6.6 GHz

2

2–

6.2 GHz to 13.2 GHz

3

4–

12.8 GHz to 19.2 GHz

4

4–

18.7 GHz to 26.8 GHz

5

4+

26.4 GHz to 31.15 GHz

6

8–

31.0 GHz to 50.0 GHz

Frequency reference

Accuracy

±[(time since last adjustment x aging rate) + temperature stability + calibration accuracy]

Aging rate

± 1 x 10

–7

/ year

Temperature stability

20 °C to 30 °C ±1 x 10

–8

0 °C to 55 °C ±5 x 10

–8

Achievable initial calibration accuracy

±7 x 10

–8

Example frequency reference accuracy
1 year after last adjustment

= ±(1 x 1 x 10

–7

+ 1 x 10

–8

+ 7 x 10

–8

)

= ±1.8 x 10

–7

Frequency readout accuracy (start, stop, center, marker)

± (marker frequency x frequency reference accuracy + 0.25% x span + 5% x RBW + 2 Hz + 0.5 x horizontal resolution*)

* Horizontal resolution is span/(sweep points – 1)

Marker frequency counter

Accuracy

±(marker frequency x frequency reference accuracy + 0.100 Hz)

Delta counter accuracy

±(delta frequency x frequency reference accuracy + 0.141 Hz)

Counter resolution

0.001 Hz

Frequency span (FFT and swept mode)

Range

0 Hz (zero span), 10 Hz to maximum frequency of model

Resolution 2

Hz

Accuracy

±[0.2% x span + span / (sweep points – 1)]