Cmu-b52 speech codec option, Spectrum due to switching – Atec R-S-CMU200 User Manual
Page 17

Version 09.00, October 2008
R&S
®
CMU200 Universal Radio Communication Tester
17
Spectrum due to switching
Reference level for full dynamic range
GMSK, RF Attenuation
→ Low Noise
RF 1
+10 dBm to +47 dBm
RF 2
–4 dBm to +33 dBm
RF 4 IN
–22 dBm to 0 dBm
Test method
absolute measurement, max. hold over
several measurements
Filter bandwidth
30 kHz resolution filter, 5 poles
Measurement
at an offset of ±
400/600/1200/1800 kHz
Dynamic range
with offset
≥ 1200 kHz
>72 dB
R&S
®
CMU-B52 speech codec option
Speech decoder output
SPEECH HANDSET OUT
9-pin D-Sub connector
Output impedance
<10
Ω
Max. output current
20 mA, peak
Full-range output level
1 V, peak
Speech coder input
SPEECH HANDSET IN
9-pin D-Sub connector
Input impedance
100 k
Ω
Full-range input level
low sensitivity
high sensitivity
1.4 V, peak
0.1 V, peak