Modulation analyzer
Level range
modulated signal
f < 2200 MHz
–40 dBm to +44 dBm
RF 1
f
≥ 2200 MHz
–34 dBm to +44 dBm
f < 2200 MHz
–54 dBm to +30 dBm
RF 2
f
≥ 2200 MHz
–48 dBm to +30 dBm
f < 2200 MHz
–80 dBm to –9 dBm
RF 4 IN
f
≥ 2200 MHz
–74 dBm to –9 dBm
Waveform quality (
ρ) uncertainty
for
ρ 0.9 to 1
<0.003
Frequency measurement range
–3 kHz to +3 kHz
Frequency measurement uncertainty
<10 Hz + drift of timebase,
see base unit specifications
Relative measurement uncertainty
result > –33 dB
<0.1 dB
Measurements
Modulation
overview
EVM versus time, graphical
ME versus time, graphical
PE versus time, graphical
I/Q analyzer
Power
standby/access-probe
power
open loop time response
narrowband power
Spectrum
adjacent channel power
Code domain power
code domain power
code domain error power
channel power
Receiver
in signaling mode
FTAP/RTAP, FETAP/RETAP