JD725A CABLE AND ANTENNA ANALYZER
Main Functions
VSWR / Return Loss
VSWR and Return Loss measurements provide the impedance per-
formance and signal reflection characteristics of cable, connectors, and
antenna systems.
Frequency range: 25 to 4000MHz
•
Dynamic Range: 60dB
•
Over 80 wireless frequency bands built-in in the instrument’s data-
•
base
Flexibility to incorporate additional frequency bands
•
User definable limit line for
•
DTF (Distance to Fault)
The DTF measurement function allows user to accurately identify faulty
locations.
Frequency range: 25 to 4000MHz
•
Distance: Up to 1250m (4125ft)
•
Dynamic Range: 60 dB
•
High resolution mode with 1001 points
•
Over 95 cable types built-in in the instrument’s database
•
Flexibility to incorporate additional cable types
•
User definable limit line for fast Pass/Fail characterization
•