Magnetic Field Probes
Probe Type
Frequency Range
Measurement
Range
Frequency Sensitivity
Type 13C
2244/90.51
3 kHz to 3 MHz
0.25 to 250 A/m
±0.5 dB (10 kHz to 3 MHz)
Type 12C
2244/90.29
300 kHz to 30 MHz
0.017 to 17 A/m
± 0.5 dB (100 kHz to 30 MHz)
Type 10C
2244/90.27
27 MHz to 1 GHz
0.025 to 16 A/m
±0.5 dB (10 to 300 MHz)
±0.65 dB (300 to 750 MHz)
±1.2 dB (750 MHz to 1 GHz)
Type 14C
2244/90.53
100 MHz to 1 GHz
0.008 to 5 A/m
±0.4 dB (100 to 300 MHz)
±0.65 dB (300 to 750 MHz)
±1.2 dB (750 MHz to 1 GHz)