Page9

Operating Area

PLZ153W

PLZ303W

PLZ603W

PLZ1003W

Operating

voltage (DC)

1.5 to 120 V

Current

30 A

60 A

120 A

200 A

Power

150 W

300 W

600 W

1000 W

Minimum operation-

starting voltage*1

0.3 V

*1 Current can flow in a range of 0.3 to 1.5 V. However, the specifications for this

instrument may not be met.

Constant Resistance Mode

PLZ153W

PLZ303W

PLZ603W

PLZ1003W

Operating range

0.1 to 10

0.05 to 5

0.025 to 2.5

0.015 to 1

Range H*4

10 to 0.1 S

20 to 0.2 S

40 to 0.4 S

66 to 1 S

1 to 100

0.5 to 50

0.25 to 25

0.15 to 10

Range L*4

1 to 0.01 S

2 to 0.02 S

4 to 0.04 S

6.6 to 0.1 S

Setting resolution

Range H*5

0.25 mS

0.5 mS

1 mS

2.5 mS

Range L*5

0.025 mS

0.05 mS

0.1 mS

0.25 mS

Setting accuracy

(current conversion)*1

±

(1% of rated current +

α

)*3

Stability

DC input voltage regulation*2

6%

±

(1000 ppm/

°

C + 1 m

/

°

C)

Temperature coefficient

at minimum resistance value

Ripple noise

Complies with graph 1 (typical values)

*1 In a range of 23

±

5

°

C and at 3 V input voltage

*2 With respect to an input voltage change of 1.5 V to 3.0 V at the minimum resistance

value. For other resistance, the voltage variations will be within 6% of the maximum

conductance (S) value of the setting range with respect to variations in all input volt-

ages. (These values were obtained with the load terminals remote sensed.)

*3

α

= 3 V

÷

set resistance - 3 V

÷

(set resistance + 8 m

)

*4 S:siemens

*5 mS: millisiemens

Note: Conductance (S) x input voltage (V) = load current (A)

Conductance (S) = 1/resistance (

)

Constant Current Mode

PLZ153W

PLZ303W

PLZ603W

PLZ1003W

Operating range

Range H

0 to 30A

0 to 60A

0 to 120 A

0 to 200 A

Range L

0 to 3 A

0 to 6 A

0 to 12 A

0 to 20 A

Setting accuracy

(with respect to rated

±

(0.3%

±

(0.3%

±

(0.3%

±

(0.3%

current)*1

+ 30 mA)

+ 60 mA)

+ 120 mA)

+200 mA)

Setting resolution

Range H

8 mA

15 mA

30 mA

60 mA

Range L

0.8 mA

1.5 mA

3 mA

6 mA

Stability

Line regulation*2

3 mA

DC input voltage

regulation*3

10 mA

Temperature coefficient

±

100 ppm/

°

C of rated current (typical value)

Ripple noise*4

RMS*5

3 mA

5 mA

10 mA

20 mA

Measuring current*6

At 30 A

At 60 A

At 100 A

At 100 A

Peak-to-peak

30 mA

30 mA

50 mA

100 mA

Measuring current

At 30 A

At 60 A

At 100 A

At 100 A

*1 In a range of 23

±

5

°

C

*2 With respect to a variation of

±

10% of the center value of input supply voltage at the

rated current of 5 V input voltage

*3 Value obtained when input voltage is varied from 1.5 V to 120 V at the current of rated

power/120 V

*4 At the rated current of 1.5 V input voltage

*5 5 Hz to 500 kHz

*6 DC to 15 MHz

PLZ-3W/3WH series

ELECTRONIC LOAD (CC·CR·CV·CP)

Specifications(PLZ-3W series)

Constant Voltage Mode

PLZ153W

PLZ303W

PLZ603W

PLZ1003W

Operating range (DC)

1.5 V to 120 V

Setting accuracy*1

±

0.1% of rated voltage

Setting resolution

30 mV

DC input current

regulation*2

±

0.01% of rated voltage

Temperature

coefficient

±

100 ppm/

°

C of rated voltage (typical value)

*1 In a range of 23

±

5

°

C (with load terminal remote sensing)

*2 With respect to a change in the input current of 10% to 100% of the rated current at 1.5

V input voltage (with load terminal remote sensing)

Constant Power Mode

PLZ153W

PLZ303W

PLZ603W

PLZ1003W

Operating range (DC)

15 to 150 W

30 to 300 W

60 to 600 W 100 to 1000 W

Setting accuracy*1

±

2% of rated power

Setting resolution

0.025% of rated power

DC input voltage

regulation*2

2% of rated power

Ripple current*3

Complies with graph 2 (typical values)

Temperature

coefficient

±

1000 ppm/

°

C of rated power (typical value)

*1 In a range of 23

±

5

°

C and at 5 V input voltage (with load terminal remote sensing)

*2 With respect to an input voltage change of 6 V to 120 V at rated power (with load

terminal remote sensing)

*3 When the ripple noise of input voltage is 5 mV or less

Tr/Tf, Switching and Soft-start Operations, Remote Sensing, and Protective Features

PLZ153W

PLZ303W

PLZ603W

PLZ1003W

Tr/Tf setting*1

Operation mode

Constant current

Setting range 50, 100, 200, 500

µ

s

1, 2, 5, 10 ms

Setting accuracy*2

±

30% of set value,

±

15

µ

s

Switching operation

Operation mode

Constant current, constant resistance

Time setting range

1 to 5000 ms

Time setting accuracy

±

5% of set value

Soft-start operation

Operation mode

Constant current(C.C)

Setting range

0.1, 1, 2, 5, 10, 20, 50, 100 ms

Setting accuracy

±

30% of set value,

±

100

µ

s

Remote sensing

Sensing voltage

5 V*3

Protective features

Over current protection

(OCP)

Limit is activated at about +5% of rated current.

Overheat protection

Load current is cut off at a heat sink

(OHP)

temperature of about 105

°

C.

Reverse connection

protection

With diodes and fuses

Power transistor

protection

With fuses

Overvoltage protection

(OVP)

Load switch will be turned off.

*1 In an input voltage range of 3 V to 120 V.

Rise time (Tr) and fall time (Tf) are times required to reach 10% to 90% of current

waveform.

*2 Tr/Tf setting is valid when the range of changes in load current is 2% to 100% of the

rated current value.

*3 2.5 V at one terminal

Advertising