2 Fluke Calibration 9190A Ultra-Cool Field Metrology Well

Wide temperature range

From –95 °C to 140 °C

Excellent accuracy
Accuracy using built-in refer-

ence thermometer readout:

± 0.05 °C full range
Display accuracy: ± 0.2 °C full

range

Best-in-class stability

± 0.015 °C full range

Fast cooling time

23 °C to –90 °C: 80 minutes

23 °C to –95 °C: 90 minutes

140 °C to 23 °C: 60 minutes

Portability
Weighs only 16 kg (35 lb)
Built-in front and back handles

for easy two-handed carry

Best measurement

practices
Conforms to EURAMET cg-13

guidance on measurement

practices for temperature block

calibrators

Ergonomics

• Two-handed carry; built-in

handles on front and back

• All controls and inputs on the

front panel

Stability indicator

• Shows whether block tem-

perature is stable and within

limits defined by the user

PC interface

• USB and RS-232 support

remote operation from a

computer

Dual heater zone

• Active heater zone control

compensates for differential

temperatures between zones

• Minimizes axial temperature

gradients

Ramp-and-soak set-points

• Define up to eight set-point

temperatures; automatically

ramp to and “soak” at each

set-point

• Specify soak time, number of

cycles, and direction

Fast time to temperature and

small footprint

• Compact, efficient free piston

Stirling cooler technology

(FPSC)

Optional process

features

Reference sensor control

• Set-point temperature control

gets passed from the

internal sensor to external

reference PRT

• 9190A controls the well tem-

perature based on its

measurement and position

inside the well

• Helps minimize the effects of

the axial gradient when

reference PRT is aligned with

short sensors

4-20 mA connectors

• Power a 4-20 mA transmitter

from the drywell

• Supplies 24 V dc loop power
4-wire PRT/RTD input

• Reads a 4-wire, 3-wire, or

2-wire PRT/RTD

• Accuracy ± 0.02 °C

• 4-20 mA circuit fuses
Thermocouple input

• Accepts common thermo-

couples with mini-jack

termination

• Types: J, K, T, E, R, S, M, L, U,

N, C

Reference thermometer input

• Smart connector input accepts

–A terminated reference PRTs

• Accuracy ± 0.01 °C at –95 °C

Features at a glance