Atec Fluke-9190A User Manual

Page 4

Advertising
background image

4 Fluke Calibration 9190A Ultra-Cool Field Metrology Well

-P Specifications
Built-in reference thermometer readout

accuracy (4-wire reference probe)

[3]

± 0.010 °C at –95 °C
± 0.013 °C at –25 °C
± 0.015 °C at 0 °C
± 0.020 °C at 50 °C
± 0.025 °C at 140 °C

Reference resistance range

0 Ω to 400 Ω

Reference resistance accuracy

[4]

0 Ω to 42 Ω: ± 0.0025 Ω

42 Ω to 400 Ω: ± 60 ppm of reading

Reference characterizations

ITS-90, CVD, IEC-751, resistance

Reference measurement capability

4 wire

Reference probe connection

6-pin din with INFO-CON technology

Built-in RTD thermometer readout accuracy

NI-120: ± 0.015 °C at 0 °C
PT-100 (385): ± 0.02 °C at 0 °C
PT-100 (3926): ± 0.02 °C at 0 °C
PT-100 (JIS): ± 0.02 °C at 0 °C

RTD resistance range

0 Ω to 400 Ω

Resistance accuracy

[4]

0 Ω to 25 Ω: ± 0.002 Ω
25 Ω to 400 Ω: ± 80 ppm of reading

RTD characterizations

PT-100 (385),(JIS),(3926), NI-120, resistance

RTD measurement capability

2-wire, 3-wire, and 4-wire RTD with jumpers only

RTD connection

4-terminal input

Built-in TC thermometer readout accuracy

[5]

Type J: ± 0.70 °C at 140 °C
Type K: ± 0.75 °C at 140 °C
Type T: ± 0.60 °C at 140 °C
Type E: ± 0.60 °C at 140 °C
Type R: ± 1.60 °C at 140 °C
Type S: ± 1.60 °C at 140 °C
Type M: ± 0.65 °C at 140 °C
Type L: ± 0.65 °C at 140 °C
Type U: ± 0.70 °C at 140 °C
Type N: ± 0.75 °C at 140 °C
Type C: ± 1.00 °C at 140 °C

TC millivolt range

–10 mV to 75 mV

Voltage accuracy

0.025 % of reading +0.01 mV

Internal cold junction compensation accuracy

± 0.35 °C (ambient of 13 °C to 33 °C)

TC connection

Miniature connectors (ASTM E1684)

Built-in mA readout accuracy

0.02 % of reading + 0.002 mA

mA range

Cal 4-22 mA, Spec 4-24 mA

mA connection

2 terminal input

Loop power function

24 V dc loop power

Built-in electronics temperature coefficient

(0 °C to 13 °C, 33 °C to 50 °C)

± 0.005 % of range per °C

Notes:

[1]

For ambient temperature of 23 °C.

[2]

Time from when the SETPOINT is reached to when the unit is with in stability specification.

[3]

The temperature range may be limited by the reference probe connected to the readout.

The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe

characterization errors.

[4]

Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05

Ω to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.

[5]

The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.