Dc power specification summary – Atec Fluke-5500A User Manual
Page 15

14
Fluke Corporation
5500A Black and White Extended Specifications
DC Power Specification Summary
Absolute Uncertainty, tcal ± 5
°C, ± (% of Watts output)
[1]
5500A Calibrator Current Range
Voltage Range
3.3 to 8.999 mA
9 to 32.999 mA
33 to 89.99 mA
90 to 329.99 mA
90 days
33 mV to 1020 V
0.03
0.02
0.03
0.02
1 year
33 mV to 1020 V
0.04
0.03
0.04
0.03
Voltage Range
0.33 to 0.8999 A
0.9 to 2.1999 A
2.2 to 4.4999 A
4.5 to 11 A
90 days
33 mV to 1020 V
0.07
0.05
0.08
0.06
1 year
33 mV to 1020 V
0.08
0.06
0.12
0.09
5725A Amplifier Current Range
Voltage Range
1.5 to 4.4999 A
4.5 to 11 A
90 days
33 mV to 1020 V
0.09
0.07
1 year
33 mV to 1020 V
0.10
0.08
[1]
To determine dc power uncertainty with more precision, see the individual “DC Voltage Specifications” and “DC Current