24
Fluke Corporation
5500A Black and White Extended Specifications
AC Voltage, Square Wave Characteristics
Rise Time
@ 1 kHz
Typical
Settling Time
@ 1 kHz
Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range
Duty Cycle Uncertainty
[1]
<1
µs
<10
µs to 1 %
of final value
<2 %
1 % to 99 %, <3.3 V p-p,
0.01 Hz to 100 kHz
± (0.8 % of period +140 ns) for
frequencies >10 kHz; + (0.8 % of
period + 2
µs) for frequencies
≤10 kHz.
[1]
For duty cycles of 10.00 % to 90.00 %.
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz
Aberrations
0.3 % of p-p value, from 10 % to 90 % point
<1 % of p-p value, with amplitude >50 % of range
AC Current (Sine Wave) Extended Bandwidth Specifications
1-Year Absolute Uncertainty,
tcal ± 5
°C,
± (% of output + % of range)
Range Frequency
% Output
% Range
Maximum
Current Resolution
All current ranges, <330 mA
0.01 to 10 Hz
5.0
0.5
2 digits each range
10 to 10 kHz
(See AC Current (Sine Wave) Specifications)