28

Command code &

Return value

Function

Example

Input command:

GETD[CR]

Return value:

[voltage][current][stat

us][CR]

OK[CR]

Get display voltage,

current, and status

reading from power

supply

[voltage] = 0000-XXXX

[current] = 0000-XXXX

[status] = 0|1 (0=CV,

1=CC)

Input command:

GETD[CR]

Return value:

030201450

OK

Indicates voltage reading is 3.02

V and current reading is 1.45 A.

Power supply is in CV mode

Input command:

GETM[CR]

Return value:

[preset 1 voltage]

[preset 1 current][CR]

[preset 2

voltage][preset 2

current][CR]

[preset 3

voltage][preset 3

current][CR]

OK[CR]

Get preset memory

values

<preset 1|2|3

voltage> = 000-XXX

<preset 1|2|3

current> = 000-XXX

Input command:

GETM[CR]

Return value:

015015

025025

035035

OK

Indicates preset 1 voltage and

current is 1.5 V and 1.5 A, preset

2 voltage and current is 2.5 V

and 2.5 A, preset 3 voltage and

current is 3.5 V and 3.5 A

Input command:

RUNM{<memory>}[CR

]

Return value:

OK[CR]

Set voltage and

current using values

saved in preset

memory locations

<memory>=0|1|2

0 = preset 1 values

1 = preset 2 values

2 = preset 3 values

Input command:

RUNM0[CR]

Return value:

OK[CR]

Sets voltage and current using

values from memory location 1