SETTING CURRENT LIMIT

1. Determine the maximum safe current for the device

to be powered. If that value is greater than 0.25 A
for Model 1710A, or greater than 0.5 A for Model
1730A, set the High-Low switch to High. If less
than these values, set High-Low switch to Low.


2. Temporarily short the (+) and (-) terminals of the

power supply together with a test lead.


3. Rotate the Coarse VOLTAGE control away from

zero sufficiently for the CC indicator to light.


4. Adjust the CURRENT control for the desired

current limit. Read the current value on the A meter.


5. The current limit (overload protection) has now

been preset. Do not change the CURRENT control
setting after this step.

6. Remove the short between the (+) and (-) terminals

and hook up for constant voltage operation.

Figure 5. Setting Current Limit.

OPERATING INSTRUCTIONS

Temporarily short

(+) to ( -)

Midrange

Adjust to

desired current

limit


CC Indicator on


Select High

Or Low

range