Full voltage calibration errors:
Er 30: CC not high
Er 31: CV not low
Er 32: DAC out of range
Er 33: ADC out of range
Er 34: ADC system calibration failed
Er 35: DAC full voltage constant checksum failed
Er 36: ADC full voltage constant checksum failed
Zero current calibration errors:
Er 40: CC not low
Er 41: CV not high
Er 42: DAC out of range
Er 43: ADC out of range
Er 44: ADC system calibration failed
Er 45: DAC zero current constant checksum failed
Er 46: ADC zero current constant checksum failed
Full current calibration errors:
Er 50: CC not low
Er 51: CV not high
Er 52: DAC out of range
Er 53: ADC out of range
Er 54: ADC system calibration failed
Er 55: DAC full current constant checksum failed
Er 56: ADC full current constant checksum failed
7.3 Command Errors
”Communication Error<CR>” : RS232 framing, parity, or overrun error
“Syntax Error<CR>” : invalid syntax was found in the command string
“Out of range<CR>” : a numeric parameter value is outside the valid range
for the command