2.1 Introduction


This section contains installation information, power requirements, initial inspection and signal
connections for the 4007B and 4013B signal generators.

2.2 Package Contents


Please inspect the instrument mechanically and electrically upon receiving it. Unpack all items
from the shipping carton, and check for any obvious signs of physical damage that may have
occurred during transportation. Report any damage to the shipping agent immediately. Save
the original packing carton for possible future reshipment. Every generator is shipped with the
following contents:

4007B / 4013B DDS function generator

AC Power Cord

Manual contained on CD

Certificate of Calibration

Verify that all items above are included in the shipping container. If anything is missing,
please contact B&K Precision.

2.4 Instrument Mounting


The 4007B and 4013B Function Generators are intended for bench use. The instrument
includes a front feet tilt mechanism for optimum panel viewing angle. The instrument does not
require special cooling when operated within conventional temperature limits. It may be
installed in a closed rack or test station if proper air flow can assure removing about 15 W of
power dissipation.

2.5 Power Requirements

The 4007B and 4013B can be operated from any source of 90V to 264V AC, frequency from
48Hz to 66Hz. The maximum power consumption is 25 VA. Use a slow blow fuse of 1A,
UL/CSA approved as indicated on the rear panel of the instrument.

The instrument power fuse is located in the AC input plug. To access the fuse, first

disconnect the power cord and then remove the fuse cartridge.