B&K Precision 4017A - Manual User Manual

Page 10

Advertising
background image

9

Figure 1. Model 4017A Front Panel.

OPERATING INSTRUCTIONS








Figure 2. Output Waveform and Phase

Relationship

The B+K Precision Model 4017A Sweep/Function Generator is a

versatile instrument, capable of producing a variety of output
waveforms over a broad range of frequencies. To gain a working
familiarity with the unit, it is recommended that it be connected
initially to an oscilloscope, so that the effects of the various controls
on the output waveforms can be observed. Use this manual as
required for reference until becoming accustomed to the operating
procedures.

FREQUENCY AND WAVEFORM SELECTION

1. Initially, verify that the DUTY CYCLE (17), CMOS LEVEL

(15), DC OFFSET (14), -20 dB (11), and SWEEP INT/EXT
(12) switches are in the OUT position (released). This will
produce a symmetrical waveform unaffected by the sweep
generator and other controls.

2. Plug the unit into an appropriate power source and turn it on by

engaging the POWER switch (1).

3. Select the desired waveform (SINE, SQUARE, or TRIANGLE)

by engaging one of the FUNCTION switches (3). Phase
relationships of the waveforms are shown in Fig. 2.

4. Select the frequency of the waveform by depressing one of the

RANGE switches (2). The output frequency is displayed, along
with the appropriate measurement units, KHz or Hz (23), on the
LED display.

5. Rotate the COARSE (20) frequency control to quickly set the

output frequency to the approximate desired value. The FINE
(19) frequency control can then be used to easily set the output
to the specific desired value. The frequency selected is available
at the OUTPUT jack (6). In addition, a digital signal, either TTL
or CMOS is available at the TTL/CMOS jack (7) (refer to the
“TTL/CMOS OUTPUT” section of this manual).

6. Adjust the amplitude of the output as desired using

the OUTPUT LEVEL control (4). Rotation of this
control varies the amplitude from maximum to 20 dB
below maximum. An additional attenuation of -20 dB
is available by engaging the -20 dB switch (11). The
attenuation factors can be combined for a total of -40
dB. The maximum signal level is 10 V p-p (into 50
Ω).

Advertising