Multimeter and recorder specifications, Included accessories – B&K Precision 2510 Series - Datasheet User Manual

Page 7

Advertising
background image

www.bkprecision.com

Handheld Digital Storage Oscilloscopes

2510 Series

7

Multimeter and Recorder Specifications

All specifications are based on operating at temperatures 23 ± 5°C and relative humidity < 75%.

Accuracy is based on ± (% of reading + offset).

Multimeter

Display Resolution

6000 counts

Measurement Function

DC voltage, AC voltage, resistance,

diode, continuity, capacitance,

DC current, AC current

Max. Input Voltage

AC: 750 V (20 Hz - 1 kHz)

DC: 1000 V

Max. Input Current

(1)

AC: 10 A (20 Hz - 1 kHz)

DC: 10 A

Input Impedance

10 MΩ

Max. Input Voltage

Between Multimeter

Input Reference

and Ground

CAT II 600 V

CAT III 300 V

DC Voltage

Range

Resolution

Accuracy

60.00 mV

10 µV

± (1 % + 15 digits)

600.0 mV

100 µV

± (1 % + 5 digits)

6.000 V

1 mV

60.00 V

10 mV

600.0 V

100 mV

1000 V

1 V

AC Voltage

(2)

Range

Resolution

Accuracy

60.00 mV

10 µV

± (1 % + 15 digits)

600.0 mV

100 µV

± (1 % + 5 digits)

6.000 V

1 mV

60.00 V

10 mV

600.0 V

100 mV

750 V

1 V

Multimeter (cont.)

Resistance

Range

Resolution

Accuracy

600.0 Ω

0.1 Ω

± (1 % + 5 digits)

6.000 kΩ

1 Ω

60.00 kΩ

10 Ω

600.0 kΩ

100 Ω

6.000 MΩ

1 kΩ

60.00 MΩ

10 kΩ

Diode and Continuity Measure

Diode

0 – 2 V

Continuity

< 50 Ω alarm

Capacitance

Range

Resolution

Accuracy

40.00 nF

10 pF

± (3 % + 10 digits)

400.0 nF

100 pF

± (4 % + 5 digits)

4.000 µF

1 nF

40.00 µF

10 nF

400.0 µF

100 nF

DC and AC Current

(3)(4)

Range

Resolution

Accuracy

60.00 mA

10 µA

± (1 % + 5 digits)

600.0 mA

100 µA

6.000 A

1 mA

± (1.5 % + 5 digits)

10.00 A

10 mA

Recorder

Scope Trend Plot

Display Mode

Full view, Normal

Record Length

800k points, > 24 hours

Number of Channels

2

Multimeter Trend Plot

Display Mode

Full view, Normal

Record Length

1.2M dots, > 24 hours

Number of Channels

1

Scope Recorder

Display Mode

Full view, Normal

Max. Record Length

Single Channel: 7 M pts

Dual Channel: 3.5 M pts

Number of Channels

2

Maximum Record Size to

External Storage

4 GB, 3000 hours

(1) Current input terminals protected with internal 250 V rated fuse.
(2) For frequency range 20 Hz to 1 kHz.
(3) For 10 A terminal, > 6 A DC or AC rms for 10 seconds ON and 15 minutes OFF.
(4) For AC current ranges, frequency is verified for 20 Hz to 1 kHz.

v040714

Included Accessories

User manual, passive probes (one per channel), pair of DMM test leads, 7.4 V Li-ion
battery BP2510, USB cable, probe compensation connector, AC power adapter,
travel hard case LC2510 (models 2515 and 2516 only), and certificate of calibration

Advertising