6 - Set-up

6.1 - Autodiagnostics

Performs tests useful for maintenance staff.

6.2 - Autocalibration

For autocalibration proceed as follows:

- Fit a wheel with steel rim of average dimensions on the shaft. Example 6” x 14” (± 1”)

- Set the exact dimensions of the wheel mounted.

CAUTION !!

Setting incorrect dimensions will result in the machine not being properly calibrated and hence all

the subsequent measurements will be incorrect until a new autocalibration is performed with the

correct dimensions!!

- Perform a manual spin

- Add a 60 g. sample weight (2.00 oz) on the outside in any angular position.

- Perform a manual spin

- Shift the sample weight from the outside to the inside keeping the same

angular position.

- Perform a manual spin

- Turn the wheel until the sample weight is in the 12 o’clock position.

END OF AUTOCALIBRATION

CANCEL AUTOCALIBRATION IN ANY PHASE.

6.2.1 – Specific procedure for exclusive use of the driving wheels (from memory M2691)

- Use one driving wheel with a rim width of not less than 5.5”
- Apply the sample adhesive weight of 60 g alternatively on the balconies as shown in the attached Figure 8a
- Measure the dimensions as indicated in Fig. 5
- Proceed with calibration as already described.

N.B.: The machine calibrated in this way, operates properly only for the correction modes shown below.