CEMB USA T5-LVDT_100-24 User Manual

Page 2

Advertising
background image

2

TECHNICAL CHARACTERISTICS

Range of measurement

: 0 to 100 mm

Composition

: Sensor T5-LVDT/100
Adapter AD1-BOX

Sensitivity

: 0,1 V/mm

Transformer LVDT

: . Impedance, primary winding = 535 ohm/2500 Hz

. Impedance, secondary windings = 875 ohm/2500 Hz
. Max linearity error = ± 1.0% f.s.
. Nominal sensitivity = 18 mV/0.001”/V
. Frequency range = 50 Hz to 10 kHz
. Supply voltage = 3 V rms (rated)

Range of application : . Temperature = -55°C to +95°C
. Humidity = max 95% not condensing
. Vibrations = max 20 g / 2 KHz
. Ambient = IP55 EN 60529/10.91 standards

Maintenance

: none

Power supply

: . Transducer = +15 Vdc, GND
Adapter = +15 to +24 Vdc, GND

Connection

: . Transducer = connector 7-pin MIL C-5015

series 3106/16, standard kit

. Adapter = connector 5-pin MIL C-5015

series 3106/16, standard kit

Outer casing material

: Aluminium

Weight

: Kg 20 approx.

Drawings enclosed

: 51351 = Overall dimensions, mounting, connections

N.B.
The sensor part carries a removable cover at the top. After removing this cover, it is
possible to read the mechanical value of turbine casing expansion on a graduated scale.