Part 3 - startup and configuration – Dynasonics 901 Series Enhanced Ultrasonic Flow Meter User Manual

Page 27

Advertising
background image


Rev.

04/02

-26-

D901/M


By applying a CAL value other than 100%, the factory
calibrated readings will be altered by the percentage
entered. This CAL value will not be reflected in the 4-
20mA or pulse outputs.

For example, if a reading of 175 GPM is displayed and the
known flow rate is 160 GPM, a CAL value of

160 GPM X 100 = 91.4%
175 GPM

The D901/M will not allow decimal values to be entered as
a CAL constant, so round to the nearest whole number. In
this case 91%.

Acceptable input ranges for the CAL constant are shown
in Table 8.









Values outside of this range will result in an OVER! or
UNDER! Display. Non-whole number entries will result in
a RANGE! Display. Enter an appropriate value.

I.D.

US

METRIC

Max

200%

200%

Min

3%

3%

Table 8

PART 3 - STARTUP AND CONFIGURATION

Advertising