Part 3 - keypad configuration, Bsc menu, Basic menu – Dynasonics TFXM Series Ultrasonic Multi-Channel Flow Meter User Manual

Page 43

Advertising
background image


Rev. 8/02

-3. 5-

TFXM

PART 3 - KEYPAD CONFIGURATION

The BASIC menu contains all of the configuration parameters
necessary to make the TFXM operational.

UNITS

ENGLSH

METRIC

Installs a global measurement standard into the operation of
the instrument. The choices are either English or Metric
measurements.

??

Select ENGLSH if all configurations (pipe sizes, etc.)are to
be made in inches. Select METRIC if the meter is to be
configured in millimeters.

??

The ENGLSH/METRIC selection will also configure the
TFXM to display sound speeds in pipe materials and liquids
as either feet per second or meters per second,
respectively.

XDCR MNT -- Transducer Mounting Method

V

W

Z

Selects the mounting orientation for the transducers. The
selection of an appropriate mounting orientation is based on
pipe and liquid characteristics. See PART 2 - Transducer
installation in this manual.

Transducer
Mount

1.

BSC MENU

-- BASIC MENU

UNITS Selection

NOTE: If the UNITS entry has been changed from ENGLSH to
METRIC or from METRIC to ENGLSH, the entry must be saved
and the instrument reset (power cycled or System Reset en-
tered) in order for the TFXM to initiate the change in operating
units. Failure to save and reset the instrument will lead to im-
proper transducer spacing calculations and an instrument that
may not measure properly.

Advertising