Btpr, Op-03 specifications – A&D Super Wash Down Scales SW-150KL User Manual

Page 30

Advertising
background image

28

1. Disconnect the AC power cable from the electrical outlet. When the battery is

installed, make sure that the scale is turned OFF.

2. Loosen the M4 screws and remove the rear cover of the display.

Note: Take care not to drop the display.

3. Connect the cables to the external device through the cable gland to the terminal

blocks on the interface board.

4. Attach the connector cables (7 and 10 pins), provided with OP-03, to the

connectors on the interface board and the connectors on the main board inside the
display.

5. Secure the interface board using the two M3 x 8 screws provided with OP-03.
6. Tighten the cable glands and attach the rear cover to the display and secure it with

the screws loosened at step 2.

7. Connect the AC power cable to the electrical outlet.

8. Set the function settings “

bp5

”, “

btpr

”, “

prt

”. “

5if

” and “

aCk

” as necessary.

‰

To use the OP-03 RS-232C, the function setting of “

5if

” must be set to “

0

”.


13.2.2. OP-03 Specifications

RS-232C interface

Transmission form

Asynchronous, bi-directional, half-duplex

Data format

Baud rate: 2400, 4800, 9600 bps

Data bits:

7 bits + parity 1bit (EVEN / ODD)

or 8 bits (non parity)

Start bit:

1 bit

Stop bit:

1 bit

Code:

ASCII

Terminator:

C

R

L

F

(C

R

:0Dh, L

F

:0Ah)

LSB

0

1

2

3

4

5

MSB

6

Relay output

The maximum rating of the replay output is as follows.
‰

Maximum voltage:

50V DC

‰

Maximum current:

100 mA DC

‰

Maximum ON resistance:

8 Ω



Stop bit

1 (-15V~-5V)

0 (5V~15V)

Start bit

Data bits

Parity bit