UNDERSTANDING PODVIEW® 

Step Six 

Linear vs. Non‐Linear:  Two of the shapes (Vertical Cylinder Tank and Rectangular Tank) will always provide a 
linear output, regardless of selecting Distance or Volume.  The remaining four shapes (Vertical Cylinder Tank 
with Cone Bottom, Horizontal Cylinder Tank with End Caps, Horizontal Cylinder Tank with Spherical End Caps 
and Spherical Tank) will have a linear output when Distance is selected, but will have a non‐linear output when 
volume is selected. 
 

Vertical Cylindrical Tank 

 

In  the  above  illustration,  10”  of  liquid  will  always 
be equal to 100 gallons of liquid (1” = 10 gallons).    

Horizontal Cylinder Tank with End Caps 

 

In  the  above  illustration,  1”  of  liquid  does  not 
equal  10  gallons.    The  10”  at  the  bottom 
represents a rise of 62.8 gallons where the change 
between  10”  and  20”  represents  an  increase  of 
109.6 gallons.  

When volume is selected, the 4‐20 mA output from the sensor will be proportional to the volume of the tank, 
not the height of the tank.  This means that the current output will track the volume of the tank (in gallons or 
liters) within a non‐linear tank (Vertical Cylinder Tank with Cone Bottom, Horizontal Cylinder Tank with End 
Caps, Horizontal Cylinder Tank with Spherical End Caps or Spherical Tank).   

When connecting the 4‐20 mA output to a display, the current signal will follow the volume of the tank.  The 
display will also reflect the volume of the tank and not the height of the liquid.   

Example #1 (Volume Output): In the illustrations above, @ 20” of liquid, the display will show 200.0 gallons in 
the Vertical Cylindrical Tank.  However, in the Horizontal Cylinder Tank with End Caps, the same level of 20” 
would show 172.4 gallons. 

 

 

 

 