Brooks, Model 5851e – Brooks Instrument 5851E User Manual
Page 10
Advertising

1-4
Brooks
®
Model 5851E
Section 1 Introduction
Installation and Operation Manual
X-TMF-5851E-MFC-eng
Part Number: 541B104AAG
November, 2008
Mounting Attitude Sensitivity
±0.5% maximum full scale deviation after re-zeroing
Command Input
0-5 Vdc. Input resistance 200 k ohm
Leak Integrity
1 x 10
-9
Atm. scc/sec Helium
Control Range
50 to 1
Mechanical Connection
Refer to Figure 2-1
Electrical Connection
Card Edge, 20 terminals, gold over low stress nickel plated copper.
D-Connector, 15 pin (DA-15P)
Environmental Conditions:
Installation Category: "I"
Pollution Degree: 2
Humidity: 50-80% Relative Humidity
Meter Altitude: 2,000 meters (6,562 feet)
For Indoor Use ONLY
Mains supply voltage fluctuations not to exceed ±10% of the nominal
voltage
Advertising